1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
247
User Manual 1173.9511.02 ─ 11
specific carrier needs to be selected via the GUI or remote command. Commands
defining the carrier configuration are independent of the pre-selected carrier.
3.2.4 Combined Signaling and Measurement
The TX measurement can be used in conjunction with the 1xEV-DO signaling applica-
tion (option R&S CMW-KS88x), i.e. a connection to the AT can be set up by the signal-
ing application and the resulting reverse signal can be measured using the TX mea-
surement.
To link the measurement to the signaling application, the combined signal path sce-
nario must be activated (see
). The signal routing and analyzer
settings are then controlled by the signaling application. The TX measurement displays
the corresponding signaling settings instead of its own settings. The configurable sig-
nal routing and analyzer settings of the signaling application can be modified both via
the measurement GUI and via the GUI of the signaling application. To configure the
signal routing, analyzer and carrier settings via remote commands, the commands of
the signaling application have to be used. For a command mapping table, see
ter 3.5.6, "Combined Signal Path Commands"
To synchronize the measurement to the signaling application, select a trigger signal
provided by the signaling application as trigger source:
●
For Multi evaluation measurements the super frame trigger of the signaling applica-
tion is recommended.
●
For OLTR measurements the "Event" trigger of the signaling application has to be
used
3.2.5 Multi Evaluation List Mode
The 1xEV-DO multi evaluation list mode requires option R&S CMW-KM012. In this
mode the measurement interval is subdivided into segments, according to the expec-
ted nominal power and frequency steps of the AT under test.
The test signal that the R&S
CMW provides in list mode can be used e.g. for fast
receiver or power meter adjustments over a wide range of channels and powers.
3.2.5.1
List Mode Configuration
Each segment contains an integer number of 1xEV-DO slots and is measured at con-
stant analyzer settings (i.e. at constant expected nominal power and RF frequency).
The figure
shows a series of four segments with different lengths, powers and fre-
quencies. Orange rectangles depict measured slots.
General Description
深圳德标仪器
135-1095-0799