1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
330
User Manual 1173.9511.02 ─ 11
<SyncState>
PEND | ADJ | INV
PEND:
waiting for resource allocation, adjustment, hardware
switching ("pending")
ADJ:
all necessary adjustments finished, measurement running
("adjusted")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
<ResourceState>
QUE | ACT | INV
QUE:
measurement without resources, no results available
("queued")
ACT:
resources allocated, acquisition of results in progress but
not complete ("active")
INV:
not applicable because <MainState>: OFF or RDY ("inva-
lid")
Example:
See
Multi Evaluation: Single-Shot and Continuous Measure-
Usage:
Query only
Firmware/Software:
V1.0.3.3
Manual operation:
See
3.5.4.2
Carrier Configuration
The following commands configure the reverse link carriers to be measured.
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:SETTing
.............................................330
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:ENABle
.............................................. 331
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FOFFset
.............................................331
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:FREQuency
........................................332
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:WBFilter
............................................. 332
CONFigure:EVDO:MEAS<i>:MEValuation:CARRier:SETTing
<SetCarrier>
Selects a carrier for the following carrier settings:
●
●
●
This command is only relevant for the standalone scenario. For the combined signal
path scenario, use:
●
●
Parameters:
<SetCarrier>
Range:
0 to 2
*RST:
0
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Command Reference
135-1095-0799