1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
111
User Manual 1173.9511.02 ─ 11
Measurement Control
The parameters in the "Measurement Control" section of the "1xEV-DO RX Measure-
ment Configuration" dialog configure the scope of the measurement.
See also: "Statistical Settings" in the R&S
CMW user manual, chapter "System Over-
view"
Fig. 2-56: 1xEV-DO RX measurement configuration: measurement control
................................................................................................ 112
Forward/Reverse Link PER > Stop Condition
.............................................................112
Forward/Reverse Link PER > Max Test Packets Sent
................................................113
Forward/Reverse Link Throughput > Max Frames
......................................................113
Update Period
Evaluation and update period of the displayed results. The results of the "RX Measure-
ment" are statistical quantities, so the R&S
CMW must collect data over some period of
time to obtain meaningful results.
The update period has no influence on the speed of the measurements, however, it
determines the shortest possible measurement time. E.g. with an update period of 1
second, new measurement reports will be available in 1.0 s intervals. The first report is
acquired during a 1.0 s period after the start of the measurement. The R&S
CMW uses
this first report to determine the offset time but discards the information. The first valid
results are calculated from the second report, which arrives between 1.0 s and 2.0 s
after the start of the measurement.