CDMA2000 Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
535
User Manual 1173.9511.02 ─ 11
Depending on your hardware configuration there may be dependencies between both
parameters. Select the RF connector first. The "Converter" parameter offers only val-
ues compatible with the selected RF connector.
Remote command:
Depends on the scenario, see
Example:
External Attenuation ← RF Input (RX)
Defines the value of an external attenuation (or gain, if the value is negative) in the
input path. The power readings of the R&S
CMW are corrected by the external attenu-
ation value.
The external attenuation value is also used in the calculation of the maximum input
power that the R&S
CMW can measure.
If a correction table for frequency-dependent attenuation is active for the chosen con-
nector, then the table name and a button are displayed. Press the button to display the
table entries.
Remote command:
4.3.7.2
RF Frequency
This section configures the band class and channel/frequency for forward and reverse
links. Additionally frequency offset is set.
Fig. 4-36: RF frequency settings
Band Class, Channel, Forward/Reverse Link Frequency
Sets the RF carrier frequency of the generator. The relation between operating band,
frequency and channel number is defined by 3GPP2. For an overview of available
band classes see
To specify the RF frequencies, select a band and enter a channel number valid for this
band. The R&S
CMW calculates the resulting frequencies for the forward and reverse
signals.
GUI Reference