CDMA2000 Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
904
User Manual 1173.9511.02 ─ 11
Parameters:
<Repetition>
SINGleshot | CONTinuous
SINGleshot:
Single-shot measurement
CONTinuous:
Continuous measurement
*RST:
SING
Example:
See
OLTR: Specifying Measurement-Specific Settings
Usage:
Query only
Firmware/Software:
V3.0.10
CONFigure:CDMA:MEAS<i>:OLTR:SEQuence
<NoOfMeasSeq>
Sets/gets the number of measurement sequences within a single OLTR measurement.
Each sequence consists of a power UP or power DOWN step, followed by a power
step in the opposite direction (see
Parameters:
<NoOfMeasSeq>
Range:
1 to 5
*RST:
1
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:CDMA:MEAS<i>:OLTR:PSTep
<PowerStep>
Defines the size of the power steps, i.e. the increases and decreases in the total BSS
power during the OLTR measurement.
Parameters:
<PowerStep>
The power step is relative to the measured reference power.
Range:
0 dB to 40 dB
*RST:
20 dB
Default unit: dB
Example:
See
OLTR: Specifying Measurement-Specific Settings
Firmware/Software:
V3.0.10
Manual operation:
See
CONFigure:CDMA:MEAS<i>:OLTR:PSTep:DIRection
<PStepDirection>
Defines the direction of the first power step within an OLTR measurement. For each
subsequent power step the direction is toggled.
Parameters:
<PStepDirection>
DOWN | UP
Command Reference