1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
439
User Manual 1173.9511.02 ─ 11
<StatisticCount>
Number of evaluated valid slots.
Comma separated list of values, one per active segment.
Example:
See
Performing Single-Shot Measurements
Usage:
Query only
Firmware/Software:
V3.0.10
Options:
R&S CMW-KM012
3.5.5 OLTR Measurement Commands
The commands for the 1xEV-DO OLTR measurement are divided into the groups listed
below. The
chapter 3.5.2, "General Measurement Settings"
the OLTR measurement.
Measurement Control and States
......................................................................... 439
.....................................................................................441
......................................................................................... 445
3.5.5.1
Measurement Control and States
The following commands control the OLTR measurement and return the current mea-
surement state.
.......................................................................................439
......................................................................................... 439
........................................................................................ 439
FETCh:EVDO:MEAS<i>:OLTR:STATe?
........................................................................... 440
FETCh:EVDO:MEAS<i>:OLTR:STATe:ALL?
.................................................................... 440
INITiate:EVDO:MEAS<i>:OLTR
STOP:EVDO:MEAS<i>:OLTR
ABORt:EVDO:MEAS<i>:OLTR
Starts, stops, or aborts the measurement:
●
INITiate...
starts or restarts the measurement; the R&S
CMW enters the
"RUN" state.
●
STOP...
causes a running measurement to stop after the current evaluation
period is terminated and valid results are available; the R&S
CMW enters the
"RDY" state.
●
ABORt...
causes a running measurement to stop immediately; the R&S
CMW
enters the "OFF" state.
Use
FETCh...STATe?
to query the current measurement state.
See also: "Measurement Control" in the R&S
CMW user manual, chapter "Remote
Control"
Example:
See
Command Reference