1xEV-DO Signaling
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
226
User Manual 1173.9511.02 ─ 11
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:FLPer:MPER
<MaxPER>
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:RLPer:MPER
<MaxPER>
Defines an upper limit for the measured forward / reverse link PER (Packet Error
Ratio).
Parameters:
<MaxPER>
Range:
0 % to 100 %
*RST:
1 %
Default unit: %
Example:
See
Firmware/Software:
V2.1.20
Manual operation:
See
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:FLPer:CLEVel
<MinConfidLevel>
CONFigure:EVDO:SIGN<i>:RXQuality:LIMit:RLPer:CLEVel
<MinConfidLevel>
Defines the minimum confidence level for the forward / reverse link PER measurement.
Parameters:
<MinConfidLevel>
Range:
0 % to 100 %
*RST:
95 %
Default unit: %
Example:
See
Firmware/Software:
V2.1.20
Manual operation:
See
2.6.11.6
Additional Softkeys and Hotkeys
The following commands provide measurement settings which are only accessible via
softkey>hotkey combinations.
CONFigure:EVDO:SIGN<i>:RXQuality:CARRier:SELect
<SelectedCarrier>
Gets/sets the carrier whose results shall be retrieved in subsequent
FETch/READ/CALCulate
commands. Applies to PL subtype 3 only.
Parameters:
<SelectedCarrier>
Range:
0 to 2
*RST:
0
Example:
See
Perform Forward Link Measurements
Firmware/Software:
V2.1.20
Options:
R&S CMW-KS881
Manual operation:
See
"Select Carrier (Hotkey; Release B only)"
Command Reference