CONFigure:EVDO:SIGN<i>:SYSTem:LSEConds
<LeapSeconds>
Adjusts track of leap second correction to UTC.
Parameters:
<LeapSeconds>
Correction to the solar time
Range:
0 to 255
*RST:
0
Default unit: s
Example:
See
Sending Date and Time Information to the AT
Firmware/Software:
V3.0.20
Manual operation:
See
CONFigure:EVDO:SIGN<i>:SYSTem:SYNC
<SyncCode>
Sets/queries the Sync Code.
The Sync Code is required to synchronize the system time for
:
query
the Sync Code generated by the "Synchronization Master" (after
SUU and
set
it on the "Syncronization Slave".
Parameters:
<SyncCode>
Firmware/Software:
V3.0.10
Options:
R&S CMW-KS888
Manual operation:
See
CONFigure:EVDO:SIGN<i>:SYSTem:ATIMe
<ApplyTimeAt>
Defines when the configured time source (
on page 177) shall be applied to the SUU hosting the signaling application.
Note
that this setting is performance critical because applying the time at signal ON
will cost 3-4 seconds.
Parameters:
<ApplyTimeAt>
SUSO | EVER | NEXT
SUSO
(Signaling Unit Startup Only): the time setting will only be
applied when the SUU starts up
EVER
: the time setting will be applied at every signal ON
NEXT
: the time setting will be applied at next signal ON; note
that after the next signal ON the R&S
CMW switches back to
SUSO
*RST:
SUSO
Example:
See
Sending Date and Time Information to the AT
Firmware/Software:
V3.0.10