1xEV-DO Measurements
R&S
®
CMW-KG8xx/-KM8xx/-KS8xx
343
User Manual 1173.9511.02 ─ 11
Parameters:
<RefPowerMode>
ATPower | PPOWer
ATPower
: total channel power
PPOWer
: pilot power
*RST:
ATP
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
2.1.10
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:LOWer
<FreqOffset9>, ..., <FreqOffset0>
Defines the negative (lower) frequency offsets to be used for ACP measurements of
the selected carrier (see
). The offsets are defined relative to the analyzer frequency.
Up to 10 offsets can be defined and enabled. The offset index 0 to 9 corresponds to
the index used in manual control.
Parameters:
<FreqOffset9> ...
<FreqOffset0>
Range:
-4 MHz to 0 MHz
*RST:
Offset 9 to 0 [MHz]: -4.0, -3.5, -3.0, -2.5, -2.0, -1.6,
-1.2, -0.9, -0.885, -0.87
Default unit: MHz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Firmware/Software:
V1.0.10.1
Manual operation:
See
CONFigure:EVDO:MEAS<i>:MEValuation:ACP:FOFFsets:UPPer
<FreqOffset0>, ..., <FreqOffset9>
Defines the positive (upper) frequency offsets to be used for ACP measurements of the
selected carrier (see
). The offsets are defined relative to the analyzer frequency.
Up to 10 offsets can be defined and enabled. The offset index 0 to 9 corresponds to
the index used in manual control.
Parameters:
<FreqOffset0> ...
<FreqOffset9>
Range:
0 MHz to 4 MHz
*RST:
Offset 0 to 9 [MHz]: 0.87, 0.885, 0.9, 1.2, 1.6, 2.0,
2.5, 3.0, 3.5, 4.0
Default unit: MHz
Additional parameters: OFF | ON (disables | enables the offset)
Example:
See
Multi Evaluation: Specifying Measurement-Specific Settings
Command Reference
深圳德标仪器