Instrument Functions
NRP2
87
User Manual 1173.9140.02 ─ 07
Note:
In addition to using an internal or external trigger source, a trigger event can be
forced using the DEL/1TRIG hardkey, see
chapter 4.2.7, "Trigger sequence control"
"Internal"
The trigger event for the selected sensor is derived from the mea-
sured signal.
"External"
The trigger event for the selected sensor is derived from the trigger
bus. The trigger signal on the trigger bus is either input at the TRIG
IN / OUT2 connector at the rear of the instrument or it is provided by
another sensor.
If the trigger signal is input at the TRIG IN / OUT2 connector, the con-
nector must be configured for "Trigger In" in the "System" menu, see
chapter 4.7.2.2, "I/O 2 interface"
If the trigger signal is provided by another sensor, this sensor must be
configured as trigger master, see below, "Int. Master".
"Int. Master"
The selected power sensor is automatically configured as trigger
master and set to internal triggering. Any previously defined trigger
master is disabled and an external trigger signal at the TRIG IN /
OUT2 connector is ignored.
The trigger master is used as the external trigger source for all other
sensors.
Remote command:
Slope
Sets the polarity of the active slope of the externally or internally applied trigger signal,
see
"
Positive"
The rising edge of the trigger signal is active.
"Negative"
The falling edge of the trigger signal is active.
Remote command:
Level
Sets the trigger threshold for internal triggering. Depending on the slope polarity, a ris-
ing or a falling edge will execute a trigger event when crossing the threshold. Modifica-
tions of the level are immediately effective (see
chapter 3.5, "Dialog boxes and their
Remote command:
Page
Switches to the next page of the dialog box.
Trigger dialog box (2)
The second page covers the parameters for defining a trigger delay and a dropout time
to prevent accidental triggering.
Data acquisition and parameters