Remote Control - Commands
NRP2
295
User Manual 1173.9140.02 ─ 07
SERVice<[1]..4>:RCOunt
<nr1>
SERVice<[1]..4>:RESult
<nr1>
Allows simulated measured values to be specified for test purposes.
SERV:RCO
speci-
fies the number of
SERV:RES
pairs to follow. The first
SERV:RES
command specifies
the measured value in watts, the second
SERV:RES
command determines how often
the measured value should be obtained. The measurement sequences defined in this
way are repeated until the sensor is disconnected from the R&S
NRP2 and reconnec-
ted.
Parameters:
<nr1>
int_value
*RST:
none. The simulation mode is not exited in the case
of *RST. To again obtain real measured values,
SERV:RCO 0 must be set.
Example:
The command sequence
SERV2:RCO 2
SERV2:RES 3
SERV2:RES 2
SERV2:RES 4
SERV2:RES 1
makes sensor 2 return the measured values 3 W, 3 W, 4 W (, 3
W, 3 W, 4W, ....) from now on.
SERVice<[1]..4>:SERRors
<state>
The sensors signal an error to the R&S
NRP2 if samples are lost while a signal is sam-
pled (sampling), as may be the case with high sampling rates. In such a case, the
measurement is less accurate. Since this error is signaled as an SCPI error, smooth
operation of a remote-control program may be disrupted. For this reason, sampling
errors can be suppressed by means of
SERVice<[1]..4>:SERRors OFF
.
Parameters:
<state>
ON | OFF
*RST:
The R&S NRP2 BIOS setting is adopted.
SERVice<[1]..4>:UNLock
<password>
Before other SERVice system commands are accepted, they must be enabled with
SERV:UNL 1234
.
Parameters:
<password>
1234
SERVice
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775