Instrument Functions
NRP2
164
User Manual 1173.9140.02 ─ 07
This way, you can test all measurement paths of R&S
NRP-Zxx multi-path sensors.
Additionally you can test an attenuator in the "... with pad" setting.
During test procedure, the instrument indicates that the test is currently running.
Fig. 4-96: Sensor test running
After completion, the instrument shows a results message.
Fig. 4-97: Sensor test results message
... with pad
Performs a test measurement for a sensor with an attached attenuator. The padded
sensor must be connected to the sensor check source. With multipath sensors, only
one path can be used for the measurement. For testing all paths, remove the attenua-
tor, connect the sensor section alone and start the test by pressing "Test at Power
Ref".
Report…
Opens the dialog box displaying the results of the test measurement.
The test report shows the type and the serial number of the tested sensor and the
detailled test results.
When the dialog box opens, it shows the data of the sensor last tested. The measured
values are lost when the R&S
NRP2 is switched off.
Fig. 4-98: Sensor test report, page 1
Page 1 shows the test results for the sensor section.