Remote Control - Commands
NRP2
269
User Manual 1173.9140.02 ─ 07
Parameters:
<state>
ON | OFF
*RST:
OFF
[SENSe<[1]..4>]:CORRection:SPDevice:SELect
<NR1>
Several S-parameter tables can be stored in a sensor.
[SENSe<[1]..4>]:CORRection:SPDevice:SELect
selects one of these tables.
The
[SENSe<[1]..4>]:CORRection:SPDevice:STATe ON
command activates
measured-value correction.
Parameters:
<NR1>
int_value
*RST:
depending on sensor
[SENSe<[1]..4>]:CORRection:SPDevice:STATe
<state>
Instructs the sensor to perform a measured-value correction by means of one of the
stored S-parameter tables (S-parameter device).
Parameters:
<state>
ON | OFF
*RST:
depending on sensor
[SENSe<[1]..4>]:FREQuency[:CW]
<frequency>
[SENSe<[1]..4>]:FREQuency[:FIXed]
<frequency>
Informs the R&S
NRP2 about the frequency of the power to be measured since this
frequency is not automatically determined. The frequency is used to determine a fre-
quency-dependent correction factor for the measurement results.
Parameters:
<frequency>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: HZ
[SENSe<[1]..4>]:FUNCtion[:ON]
<mode>
Sets the sensor to the selected measurement mode.
In remote control, the timing (
SENS:TIM:EXCL:STAR
and -
:STOP)
and the trigger
system of all measurement modes (except the Burst mode) are user-selectable.
Note: