Remote Control - Commands
NRP2
270
User Manual 1173.9140.02 ─ 07
To increase measurement accuracy, the sensors measure each signal twice in the
Burst, Timeslot and Trace measurement modes (exception: in the Trace mode, with
SENS:TRAC:REAL ON
). As long as the trigger system is set to internal or external trig-
gering, this is usually of no importance to the user. If
TRIG:SOUR HOLD
or
BUS
is
used, however, then each measurement must be triggered twice (with
*TRG
or
TRIG:IMM
).
Error messages:
24
"Sensor mode not supported"
: A sensor does not support a measurement mode.
28
"Sensor not idle"
: The sensor is not in the IDLE state.
-151
"Invalid string data"
: An invalid parameter was transmitted for <string>.
SENSe (Sensor Commands)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com