Instrument Functions
NRP2
86
User Manual 1173.9140.02 ─ 07
► Configure the trigger system in the "Trigger" dialog box which is opened via the
"Sensor" menu, or via the control panel of the respective trace mode.
Trigger dialog box, page1
The first page covers the parameters for defining a trigger mode, trigger source, and
trigger level.
Fig. 4-20: Trigger dialog box, page1
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Mode
Selects the trigger mode.
Selections "Single" and "Normal" are only available in "Trace&Statistics" measurement
mode.
"Single"
"Trace&Statistics" mode only. This setting disables continuous trig-
gering so that only one trigger event at a time is executed. To enable
triggering again, press the (DEL/1TRIG) key. The symbol
appears in the display (see
chapter 4.2.7, "Trigger sequence control"
"Normal"
Continuous triggering with regular trigger events.
"Auto"
"Trace&Statistics" mode only. This setting automatically starts a mea-
surement if no trigger event has occurred after 300 ms.
Remote command:
Source
Selects the trigger source for the selected sensor. The measured internal signal ("Inter-
nal") or an external signal is available for triggering. The external trigger signal is either
input at the TRIG IN / OUT2 connector at the rear of the instrument ("External") or it is
supplied by another power sensor which must be configured as a trigger master ("Int.
Master").
Data acquisition and parameters