Remote Control - Commands
NRP2
243
User Manual 1173.9140.02 ─ 07
Parameters:
<display_mode>
DIGital | ANALog | GRAPhical
DIGital
Measured values are displayed in numeric format. If the primary
channel is operated in Trace mode, the (scalar) measured val-
ues of the currently selected timegate are displayed here.
ANALog
Measured values are indicated by a marker on a scale. The val-
ues at the upper and lower scale end as well as the type of scal-
ing (linear or logarithmic) are set with the following
:METer
com-
mands.
GRAPhical
Measured values are plotted over time. This kind of presentation
is possible only if the primary channel is operated in Trace
mode.
*RST:
DIGital
DISPlay[:WINDow]<[1]..4>:METer:AUTO
<auto>
DISPlay[:WINDow]<[1]..4>:ANALog:AUTO
<auto>
DISP:AUTO ONCE
automatically determines scaling for the analog display. The upper
and the lower limit value of the display is set as a function of the current measurement
data. Only the
ONCE
parameter is permissible; a query always supplies the value
OFF
.
Parameters:
<auto>
ONCE | OFF
*RST:
OFF
DISPlay[:WINDow]<[1]..4>:METer:LOWer
<lower_limit>
DISPlay[:WINDow]<[1]..4>:ANALog:LOWer
<lower_limit>
Sets the lower limit value of the analog scale.
The value range depends on the current output unit of the measured value of the win-
dow.
Table 6-12: Value range
Unit of measured value
W
DBM
PCT
DB
Value range for <float_value>
1e-18..1e18
-150..210
1e-18..1e22
-200..200
Unit:
Value ranges, depending on the unit
.
Parameters:
<lower_limit>
float_value
*RST:
-90 DBM or -120 DB
DISPlay