Instrument Functions
NRP2
129
User Manual 1173.9140.02 ─ 07
"Pulse Off
Time"
Indication of the gap between two pulses.
"Rise / Fall
Time"
Indication of the times the signal requires to change from the power
off to power on state.
"Start / Stop
Time"
Indication of the start / stop point of the first power pulse within the
analysis window, i.e. the time when the signal passes through the
medial reference power level with rising or falling edge, referenced to
the delayed trigger event.
Pulse Measurements / Power
Fig. 4-57: Pulse Measurements / Power tab
The pulse power parameters to be displayed in the info panel are selected on the
"Power" tab.
"Top / Base
Power"
Indication of the power of the pulse top / base, defined in the
"Average (1
period)"
Indication of the average power of the signal. The visible trace length
must exceed one cycle of the waveform. The gate length is arbitrary.
"Max (Peak)"
Indication of the maximum power measured within the analysis win-
dow.
"Minimum"
Indication of the minimum power measured within the analysis win-
dow.
"Pos. / Neg.
Overshoot"
Indicates the relative amount of positive / negative overshoot. The
indicates value depends on the "Related to" (Power / Voltage) setting
in the pulse measurement "Config" menu.
Pulse Measurements / Config
Fig. 4-58: Pulse Measurements / Config tab
Displaying traces
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com