Annex
NRP2
357
User Manual 1173.9140.02 ─ 07
Control
character
Interface function
DT1
Device trigger
C0
No controller capabilities
A.2.1.5
Interface messages
Interface messages are transmitted to the device on data lines, where the ATN is set to
active LOW. They are used for communication between the controller and the
R&S
NRP2.
Common commands
Common commands are in the code range 0x10 to 0x1F. They affect all devices on the
bus without any addressing being required.
Table 1-2: Common commands
Command
Effect on device
DCL (Device Clear)
Aborts all commands and messages in progress. The
TRIGger
sys-
tem goes to the "IDLE" state, input and output queues are cleared,
and the parser is reset and waits for the beginning of a new com-
mand. When
INITiate:CONT ON
is set, a new measurement is
started. This command makes the device ready to receive com-
mands.
IFC (Interface Clear)
Resets the interfaces to their initial condition.
LLO (Local Lockout)
Manual switchover to LOCAL is disabled.
L
ocal
L
ock
O
ut deactivates the front-panel keypad with the exception
of the power key
. In the LLO state, it is possible to return to the
"LOCAL" mode only by remote control (
GTL
command) or switching
the R&S
NRP2 off and on again.
SPE (Serial Poll Enable)
Ready for serial poll.
SPD (Serial Poll Disable)
End of serial poll.
PPU (Parallel Poll Unconfigure)
End of parallel poll query status.
Addressed commands
Addressed commands are in the code range 0x00 to 0x0F hex. They only affect devi-
ces addressed as a listener.
Table 1-3: Addressed commands
Command
Effect on device
SDC (Selected DeviceClear)
Interrupts processing of received commands and sets the command pro-
cessing software to a defined initial state. This command does not
change the device setting.
GTL (Go to Local)
Change to local mode (manual control).
PPC (Parallel Poll Configure)
Configures device for parallel poll.
Remote control interfaces