Remote Control Fundamentals
NRP2
172
User Manual 1173.9140.02 ─ 07
5.2.2 Hardware Requirements
Risk of network failure
Before connecting the instrument to the network or configuring the network, consult
your network administrator. Errors may affect the entire network.
GPIB (IEC/IEEE bus)
In order to remote control the R&S
NRP2 via the GPIB bus, a controller with the
respective bus interface and the bus cable is required.
USB
The R&S
NRP2 automatically recognizes a controller connected to the USB1.1 inter-
face.
Ethernet/ VXI11
Controlling the R&S
NRP2 via the Ethernet interface requires a computer with network
card and LAN cables to connect the computer and the R&S
NRP2 controller with the
Ethernet.
5.3 Switchover to Remote Control (REMOTE)
It is assumed that a link has been established between the controller and the
R&S
NRP2 has been correctly configured.
After power up, the R&S
NRP2 is always in manual control mode "LOCAL". The
R&S
NRP2 switches to remote control irrespective of the selected interface as soon as
it receives a SCPI command. All settings are maintained when the device switches to
remote mode, except for the setting listed in
5.4 Return to Manual Operation (LOCAL)
In the remote mode, all front-panel controls are disabled, except for the ESC/LOCAL
key and the topmost softkey. The softkey used to open the window with the SCPI error
queue. The device remains in remote mode until it is switched back to manual control
either by using the front panel ESC/LOCAL or with the remote control command
GTL
.
Switchover to Remote Control (REMOTE)