Remote Control - Commands
NRP2
329
User Manual 1173.9140.02 ─ 07
........................................................................................... 333
............................................................................................... 333
TRIGger<[1]..4>:MASTer[:STATe]
................................................................................... 334
............................................................................................... 334
TRIGger<[1]..4>:SYNChronize[:STATe]
............................................................................335
ABORt<[1]..4>
ABORt:ALL
Immediately sets the respective sensors to the IDLE state. Measurements in progress
are interrupted. If
INIT:CONT ON
is set, a new measurement is immediately started
since the trigger system is not influenced.
Usage:
Setting only
INITiate<[1]..4>:CONTinuous
<state>
INITiate:ALL:CONTinuous
<state>
Selects either single-shot or continuous (free-running) measurement cycles. In the
course of a measurement cycle, a sensor passes through the IDLE, INITIATED,
WAIT_FOR_TRG and MEASURING states before it is again set to the IDLE (
INIT:
CONT OFF
) or the INITIATED (
INIT:CONT ON)
state. The settings are not changed
when control is switched from manual to remote.
INIT:CONT ON
is set when remote
control is switched back to manual control.
Parameters:
<state>
ON | OFF
ON
Continuous measurements are performed. If a measurement is
completed, the respective sensors do not return to the IDLE
state but are immediately set to INITIATED and then to
WAIT_FOR_TRG.
OFF
A measurement cycle is only performed once. After completion,
the sensors remain in the IDLE state.
INIT:CONT OFF
has no
effect when the sensor is already in the IDLE state. A measure-
ment in progress is completed.
*RST:
OFF
Pressing the PRESET key or sending the
SYSTem:PRESet
command sets the
INITiate:ALL:CONTinuous
command to
ON
.
TRIGger