Instrument Functions
NRP2
62
User Manual 1173.9140.02 ─ 07
●
all channels in the Cont Av mode (continuous average power)
●
absolute power measurement in dBm
●
one window per channel opened
●
offset: 0 dB
●
automatic filtering (normal mode)
(PRE)SET does
not
reset the following settings:
●
offset tables
●
address of the remote interface
●
SCPI error queue
●
selection of the beeper
●
brightness setting of the display
●
test limits for sensor test
SCPI command:
on page 198 (Default state for remote-control operation)
or
on page 325 (Default state for manual operation)
4.1.2 Frequency hardkey
FREQ is used to set the carrier frequency of the applied signal. This corrects any fre-
quency-dependent effects introduced by the sensor.
If the sensor is to reach the specified measurement accuracy, the signal frequency
must be entered.
4.1.2.1
Frequency dialog box
Fig. 4-1: Frequency dialog box
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.