Instrument Functions
NRP2
71
User Manual 1173.9140.02 ─ 07
4.2.1.5
Measuring the average power in gates
The timegate mode is used to measure the average power in arbitrarily chosen time
intervals (gates). The time resolution is determined by the sampling rate of the sensor.
Up to four different gates can be defined, although measurement can be performed
only in one of them at a time. The "T'gate Av" mode is set and configured in this dialog
box. The active gate must be selected in the dialog box "Func
ter 4.4.3, "Selecting a measurement function"
The R&S
NRP-Z8x wideband power sensors can additionally define an exclusion inter-
val within the gate (Fence). A midamble of the signal can thus be excluded from the
measurement.
An external trigger signal or internal signal triggering is required for synchronization,
see also
chapter 4.2.6, "Trigger settings"
► Select "T'gate Av" in the "Mode..." dialog box opened via the "Sensor" menu.
Fig. 4-9: Mode dialog box / T'gate Av
Gate
Sets the number of the gate to be configured.
Note:
In remote control, the gate to be configured is determined by the
TGATe
suffix,
see commands for the following settings.
Start/Length of Gate
Start of the gate is referenced to the delayed trigger time, see
>
The end of the gate is determined by the start time and the length of the gate.
Remote command:
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:OFFSet:TIME
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:TIME
Fence
Activates the "Fence" function.
In numeric T'slotAv mode, the "Fence" function is only available for the wideband
power sensors of the R&S NRP-Z8x series. For all other sensors, a fence can be used
in the "T'slots" display in the Trace&Statistics sensor mode.
Data acquisition and parameters