Instrument Functions
NRP2
104
User Manual 1173.9140.02 ─ 07
"Pk/Avg"
Ratio of peak power to average power in the gate (only in the "Trace
& Statistics" mode)
Remote command:
Selection of display
DISPlay[:WINDow]<[1]..4>:AVALue
Reading out values
4.4.3.3
Measuring power in timeslots
As with time-gated power measurements, the R&S
NRP2 power meter also offers two
measurement modes for measurements within timeslots
●
T'Slot Av
●
Trace & Statistics
T'Slot Av
In this mode, the average power (
Av
) within any timeslot of a user-definable timeslot
structure can be measured. The time resolution is high and corresponds to the sam-
pling rate of the sensor.
To configure the measurement:
1. Select "T'Slot Av" in the "Mode" dialog box of the "Sensor" menu
2. Set the timing parameter for the timeslots
3. Select the timeslot number for the measurement in the "Function & Unit" dialog box
of the "Measurement" menu.
The average power within the selected timeslot is indicated.
Fig. 4-36: Measurement within a timeslot (T'Slot Av mode)
Trace & Statistics
In this mode, the peak power (
Pk
) or the
Pk/Av
ratio can be measured within any time-
slot of a user-definable timeslot structure in addition to the average power (
Av
). How-
ever, the time resolution is limited to the length of one point (pixel) of the display.
The measured values can either be output as additional numeric information in a
graphic window or in a numeric window ("Dig", "D&A").
Configuring measurements