![Rohde & Schwarz NRP2 User Manual Download Page 135](http://html.mh-extra.com/html/rohde-and-schwarz/nrp2/nrp2_user-manual_1477858135.webp)
Instrument Functions
NRP2
132
User Manual 1173.9140.02 ─ 07
I
←
= start of measurement
→
I
= stopt of measurement
I
←→
I
= duration of measurement
Time slots are marked by dotted lines. The period of the currently selected time slot
over which the signal is averaged is marked by a colored frame with a continuous
lines. If exclude times have been set this period is shorter than the nominal time slot
width.
The defined time slot structure is common for all diplayed waveforms. Individual time
slot structures for specific channels (sensors) can only be configured in numeric "T'slo-
tAV" mode.
Tips for power measurement in timeslots
●
Load mobile radio standard (
chapter 4.6.4, "Loading settings for mobile radio
on page 142) or enter the parameters of the timeslot structure in the
"Config T'Slots" menu.
●
Set screen so that the envelope of a complete frame is displayed.
●
Open "Trigger" dialog box and set "Delay" so that the timeslot structure coincides
with the measured envelope.
●
To increase the measurement accuracy,
–
expand the window horizontally to such an extent that the selected timeslot is
displayed in adequate size
–
perform measurement in the numeric "Timeslot Av" mode.
This mode operates with precisely the time resolution available from the sensor
(sample period). In the graphic mode the time resolution is determined by the
width of a point (pixel).
SCPI commands:
Select sensor function:
Define calculation of measurement result:
Read result:
4.5.7.1
Measurement results
The measured values for the selected time slot and trace (waveform) are indicated in
the info panel below the graph.
Displaying traces