Remote Control - Commands
NRP2
200
User Manual 1173.9140.02 ─ 07
Command
Preset and *RST value
Remark
DISP<[1]..4>:MET:LOW
-90 DBM
-120 DB
Lower limit value for analog result
display.
DISP<[1]..4>:MET:UPP
70 DBM
60 DB
Upper limit value for analog result
display.
DISP<[1]..4>:NAME
""
Text to be displayed in result win-
dow.
DISP<[1]..4>:RES
0.01
Measurement accuracy and dis-
play accuracy.
DISP:SEL
1
Selection of first result window.
DISP:SIZE
NORM
Automatic window sizes.
DISP<[1]..4>
ON | OFF
A window will automatically be
opened for each sensor.
DISP<[1]..4>:TRAC:LOW
-120 DBM
Lower limit of power axis in Trace
mode
DISP<[1]..4>:TRAC:UPP
30 DBM
Upper limit of power axis in Trace
mode
DISP<[1]..4>:TSL
1
First timeslot
DISP<[1]..4>:UPD
NORM
Frequent display updates.
FORM
ASC
Measured data in ASCII format.
FORM:BORD
NORM
"Big-Endian" byte sequence of
measured data, if
FORMat REAL
is selected.
INIT[1..4|:ALL]:CONT
ON
OFF
PRES and
SYSTem:PRESet
.
*RST
INIT[1..4|:ALL]:DIS
OFF
MEM:…
-
The settings in the
MEMory
com-
mand system will not be affected.
OUTP:REC1:FEED
"CALC1"
Analog outputs are fed by the first
calculate block.
OUTP:REC2:FEED
"CALC2"
Analog outputs are fed by the sec-
ond calculate block.
OUTP:REC[1..2]:LIM:LOW
-20 DBM
or
0 DB
0 V equivalent at analog output.
OUTP:REC[1..2]:LIM:UPP
20 DBM
or
10 DB
3 V equivalent at analog output.
OUTP:REC1:STAT
ON
Analog output active.
OUTP:REC2:STAT
OFF
Trigger input active.
OUTP:ROSC
OFF
Test generator is switched off.
Common Commands to IEEE 488.2