Remote Control - Commands
NRP2
214
User Manual 1173.9140.02 ─ 07
<end_exclude>
Value range: depending on sensor
Default value: depending on sensor
Unit: S
Amount of time at the end of a timeslot or integration period that should not be
taken into account. Corresponds to command:
[SENSe<[1]..4>]:TIMing:EXCLude:STOP
<expected_value>
Value range: float_value | DEF
According to the SCPI-1999.0 standard, a value in the order of the expected
measured value can be specified. This allows the device to set an adequate
measurement range. This is not required in the R&S
NRP2 because the mea-
surement is performed simultaneously in all measurement ranges. Any value
can therefore be specified for the expected_value parameter, especially the
DEFault value. The value is always ignored.
<no_slots>
Value range: depending on sensor
Default value: depending on sensor
Unit: -
Number of timeslots to be measured in the Timeslot mode. Corresponds to
command:
[SENSe<[1]..4>][:POWer]:TSLot:COUNt
<resolution>
Value range: 1 | 0.1 | 0.01 | 0.001 |
1 | 2 | 3 | 4
Default value: 3 (0.01)
Specifies the limit up to which the measurement result should be free of noise.
The R&S
NRP2 tries to set an adequate window for the averaging filter. This
does not affect the accuracy of the measured data. The window can only be set
if automatic setting of the filter window has not been disabled by the user
(
SENSe:AVERage:COUNt:AUTO OFF
). Since the automatic setting is activated
by
CONFigure
and
MEASure?
as a default, the user normally does not need to
alter this setting. Corresponds to commands:
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:RESolution
DISPlay[:WINDow]<[1]..4>:RESolution
Example:
If
<resolution>
is set to 3 or 0.01, three significant points are assigned to the
value when it is displayed on a linear scale. When a log scale is used, 2 places
are displayed after the decimal point.
<scope_size>
Value range: depending on sensor.
Default value: depending on sensor.
Unit: -
The number of test points on the time axis. Corresponds to command:
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)