
Getting Started
NRP2
47
User Manual 1173.9140.02 ─ 07
For a detailed description of these parameters, refer to
2.8 Measuring average burst power (Burst Av mode)
An R&S
NRP-Z1x, R&S
NRP-Z2x(x), or R&S
NRP-Z8x sensor is required for this mea-
surement.
To perform a burst-average power-measurement, you need a pulsed RF signal. If you
use an R&S
NRP-Z1x or R&S
NRP-Z2x(x) sensor, the pulse repetition frequency of the
signal should not exceed 10 kHz.
Setting the burst mode.
1. Connect the power sensor to connector A on the R&S
NRP2 and apply a pulsed
signal with a level between –10
dBm and +10
dBm.
2. Press the (PRE)SET hardkey twice.
A measurement window is now available.
3. Change to the "Sensor" menu and select "Mode".
The "Mode" dialog box opens.
4. Select "Trace & Statistics".
5. In the trace mode check that the sensor detects a signal and reliably triggers.
For information concerning reliable triggering see
chapter 2.7, "Graphically repre-
6. Change back to the "Sensor" menu and select "Burst Av" in the "Mode" dialog box.
Fig. 2-38: Mode dialog box / Burst Av
The burst mode parameters are displayed at the bottom of the dialog box.
To exclude the pulse build-up and decay phases from the measurement, you can
set the "Excluded from Start" and "...from End" parameters accordingly.
Measuring average burst power (Burst Av mode)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com