Error Messages and Troubleshooting
NRP2
343
User Manual 1173.9140.02 ─ 07
Error code
Error text returned upon queue query
Explanations
-223
"Too much data"
A valid parameter is received but it contains more data than the device can handle.
-224
"Illegal parameter value"
The parameter value is invalid.
-225
Out of memory
The device has insufficient memory to perform the required operation.
-226
"Lists not same length"
A list of offset values and frequencies has not the same length as an associated list of fre-
quency and offset values.
-230
"Data corrupt or stale"
The data are incomplete or invalid.
-240
"Hardware error"
The command cannot be executed because of a hardware problem in the device.
-241
"Hardware missing"
The command cannot be executed because of missing device hardware.
8.2.4 Device-Specific Errors
Device-specific error; causes bit 5 to be set in the ESR register
Error code
Error text returned upon queue query
Explanations
-300
"Device-specific error"
Generic device-dependent error that cannot be defined more precisely.
-310
"System error"
Indicates that some system error has occurred. Please contact the R&S service center.
-311
"Memory error"
An error was detected in the device's memory.
-314
"Save/recall memory lost"
Indicates that the nonvolatile data saved by the
*SAV?
command has been lost.
-315
"Configuration memory lost"
Indicates that nonvolatile configuration data saved by the device has been lost.
-330
"Self-test failed"
The self-test could not be executed.
-341
"Zeroing failed"
The zeroing of a sensor (
CALibration:ZERO:AUTO ONCE
) has failed. The most frequent
cause for this error is a signal that is applied to the sensor input and prevents zeroing.
SCPI-Specific Error Messages
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com