Remote Control - Commands
NRP2
263
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:TRACe:MEASurement:TRANSition:POSitive:OCCurrence?
[SENSe<[1]..4>]:TRACe:MEASurement:TRANSition:POSitive:OVERshoot?
[SENSe<[1]..4>]:TRACe:MEASurement:TRANSition:SPERiod?
..........................................291
[SENSe<[1]..4>]:TRACe:MEASurement[:STATe]
...............................................................291
[SENSe<[1]..4>]:TRACe:MID:OFFSet:TIME
......................................................................292
[SENSe<[1]..4>]:TRACe:MID:TIME
..................................................................................292
[SENSe<[1]..4>]:TRACe:OFFSet:TIME
............................................................................ 292
...................................................................................... 292
[SENSe<[1]..4>]:TRACe:REALtime
..................................................................................293
.........................................................................................293
[SENSe<[1]..4>]:AUXiliary
<mode>
Activates the measurement of additional measured values that are determined
together with the main measured value.
Parameters:
<mode>
NONE
No additional values are measured.
MINMax
By averaging the measured values in the sensors, extreme val-
ues are lost. Using
[SENSe<[1]..4>]:AUXiliary MINMax
,
a sensor can be assigned to transmit to the R&S
NRP2 the min-
ima and maxima that occurred in the Trace mode together with
the measured value. If
CALCulate:FEED "POWer:PEAK ON
SWEep<[1]..4>"
is set, peak (instead of average) values are
used for calculating.
RNDMax
In contrast to
MINMax
, instead of the Min value the value of a
randomly selected sample is returned. All evaluations occur
using these values instead of the average values.
*RST:
depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt
<filter_length>
Sets the filter length i.e. the number of readings to be averaged for one measured
value.. The higher the count the lower the noise and the longer it takes to obtain a
measured value.
Parameters:
<filter_length>
int_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: -