Annex
NRP2
403
User Manual 1173.9140.02 ─ 07
E4418B/E4419B command
Compatibility information
TRIGger[:SEQUence[1]|2]
The R&S
NRP2 does not recognize the
:SEQUence
header and
uses no
TRIGger
command globally but only on a sensor-by-sensor
basis. If
TRIGger
or
SENSe
commands are to have a global effect,
the
:ALL
header must be used instead of the numeric suffix.
UNIT:POWer
The R&S
NRP2 recognizes the units
W
,
DBM
and
DBUV
.
UNIT:POWer:RATio
The R&S
NRP2 recognizes the units
DB
,
DPCT
and
O
.
:AC
Agilent recognizes the keyword
:AC
of some high-level commands
and
SENSe
commands:
CONFigure | READ | MEASure | FETCh
[:SCALar][:POWer:AC]
:RELative
:DIFFerence
:DIFFerence:RELative
:RATio
:RATio:RELative
SENSe:POWer:AC:RANGe:AUTO
The R&S
NRP2 also recognizes these commands but
:AVG
should
be used instead of
:AC
since R&S
NRP-Zxx Power Sensors sensors
eventually also measure DC components and the designation in the
R&S
NRP2 is not used in the sense of SCPI.
Compatibility Information