Remote Control - Commands
NRP2
212
User Manual 1173.9140.02 ─ 07
Usage:
Query only
6.4.3 Syntactical Structure of High-Level Measurement Commands
The high-level measurement commands can be generally represented as follows:
<head><meas_mode>[<calc_function>][?] <parameter_list>
Table 6-5: Syntactical Structure of High-Level Measurement Commands
<head>
<head> = These keywords select the high level command subsystem.
Command keywords
●
CONFigure<[1]..8>
●
FETCh<[1]..8>
●
MEASure<[1]..8>
●
READ<[1]..8>
<meas_mode>
<meas_mode> = These keywords select the measurement mode ( ContAv, ContAv
with data buffering, Burst, Timeslot, Trace, or Statistics, see also low lwvwl command
Command keywords
●
[:SCALar][:POWer][:AVG]
(ContAv mode)
●
[:SCALar][:POWer]:TSLot
(T'slot mode)
●
[:SCALar][:POWer]:BURSt
(Burst mode)
●
:ARRay[:POWer][:AVG]
(ContAv with data buffering mode)
●
:XTIMe[:POWer]
(Trace mode)
●
:STATistics:CCDF
(Statistics mode)
●
:STATistics:PDF
Statistics mode)
<calc_function>
<calc_function> = These keywords select the calculation function, see also the low
level command
CALCulate<[1]..8>:MATH[:EXPRession]
<calc_function> is not specified, the measured value of the primary channel is output.
Command keywords
●
:RELative
●
:DIFFerence
●
:DIFFerence:RELative
●
:SUM
●
:SUM:RELative
●
:RATio
●
:RATio:RELative
●
:RLOSs
●
:SWR
●
:REFLection
[?]
The "?" terminates the measurement commands which are always queries. Measure-
ment commands are the
FETCh:...?
,
READ:...?
and
MEASurement:...?
com-
mands.
<param_list>
A list with parameters is sent with the high level command. The meaning of the
parameters is determined by the
<meas_mode>
chapter 6.4.4, "Parameter list
6.4.4 Parameter list description
Six different parameter lists are used in the high end measurement commands. Which
list is used depends on the selected measurement mode. The elements of these lists
can be omitted starting at the end and working backward. They will then be replaced
by their default values.
High-Level Measurement Commands (CONFigure, MEASure, READ and FETCh)
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com