Remote Control - Commands
NRP2
318
User Manual 1173.9140.02 ─ 07
Table 6-32: Initialization of device status
Result Power On
DCL, SDC
(Device
Clear,
Selected
Device
Clear)
*RST or
SYSTem:
PRESet
STATus:
PRESet
*CLS
Power-On-
Status
Clear
(*PSC)
Effect
0
1
Clear STB (Status Byte) and
ESR (Standard Event Status Register)
--
yes
--
--
--
yes
Clear SRE (Service Request Enable)
and
ESE (Standard Event Status Enable)
--
yes
--
--
--
--
Clear PPE (Parallel Poll Enable)
--
yes
--
--
--
--
Clear EVENt register
--
yes
--
--
--
yes
Clear ENABle register of all OPERation
and
QUEStionable registers
--
yes
--
--
yes
--
Fill PTRansition register with "1",
delete NTRansition register
--
yes
--
--
yes
--
Clear error/event queue
yes
yes
--
--
--
yes
Clear output queue
yes
yes
yes
1)
1)
1)
Stop command processing (parser) and
clear input queue
yes
yes
yes
--
--
--
1):
Any command sent to a non-empty output queue causes the error:
-410
"Query interrupted"
6.14 SYSTem
With the aid of the
SYSTem
command system, administrative device settings can be
made and queried.
This includes:
●
list of error messages
●
IEC/IEEE-bus address
●
resetting the R&S
NRP2
●
setting the date and time
●
audible signals
●
version numbers of hardware and software
●
setting the system speed
SYSTem