Instrument Functions
NRP2
128
User Manual 1173.9140.02 ─ 07
Auto
Selects a vertical scaling that is suitable for the signal.
Gate
Selects if the pulse analysis is performed for the visible trace or within the selected
gate.
"[1] [2] 3 [4]"
Pulse analysis is performed within the selected gate.
"Trace"
Pulse analysis is performed for the visible trace (entire screen width).
Select Trace
Switches between the traces available for pulse analysis (available only for waveforms
representing absolute power of a wideband power sensor (R&S NRP-Z8x series)).
4.5.6.2
Pulse diagram settings
► Set the parameters of pulse analysis in the "Config Pulse..." dialog box opened via
the Window
→
Graph
→
"Display"
→
"Pulse" control panel ("Trace&Statistics" sensor
mode).
The R&S
NRP2 indicates the total number of the activated parameters in the header of
the dialog box. You can activate up to 6 items out of 15 for display.
As a prerequisite of pulse analysis, the parameters selected for display must be inclu-
ded in the analysis window. If for example the duty cycle of a pulsed signal is to be
determined, at least one cycle of the waveform must be within the borders of the
selected gate or must be part of the visible portion of the trace (if pulse analysis
extends over the whole trace). If the analysis window contains several cycles, evalua-
tion always applies to the first cycle.
Pulse Measurements / Time
Fig. 4-56: Pulse Measurements / Time tab dialog box
The pulse time parameters to be displayed in the info panel are selected on the "Time"
tab.
"Pulse Width"
Indication of pulse duration.
"Pulse Period"
Indication of the time the pulse signal needs to complete one cycle.
"Duty Cycle"
Indication of the duty cycle of the measured power.
Displaying traces