Instrument Functions
NRP2
150
User Manual 1173.9140.02 ─ 07
Table 4-12: Example of a network configuration for a direct R&S
NRP2 – PC connection using a
crossover cable.
R&S
NRP2
PC
"IP Address"
192.168.0.2
192.168.0.1
"Mask"
255.255.255.0
255.255.255.0
"Gateway"
192.168.0.3
192.168.0.3
DHCP ← IP Config
Fig. 4-78: IP Config dialog box / DHCP
Note:
For automatic assignment, a DHCP server must be available in the network.
If the R&S
NRP2 is to be integrated into a larger network, contact your network admin-
istrator regarding the data to be set.
The instrument assigns address and gateway automatically. The automatic configura-
tion mode uses DHCP server to obtain the instrument IP address.
Command Set
Selects the instrument type for emulation, see
4.7.1.5
Raw Ethernet (socket communication) settings
An alternative way for remote control of the software is to establish a simple network
communication using sockets. The socket communication, also referred as “Raw
Ethernet communication”, does not require a VISA installation on the remote controller
side.
The simplest way to establish socket communication is to use the build-in telnet pro-
gram. The telnet program is part of every operating system and supports a communi-
cation with the software on a command-by-command basis. For better utilization and to
enable automation by means of programs, user defined sockets can be programmed.
Socket connections are established on a specially defined port. The socket address is
a combination of the IP address or the host name of the instrument and the number of
the port configured for remote-control. The R&S
NRP2 uses port number 4002 for this
purpose. The port is configured for communication on a command-to-command basis
and for remote control from a program.
For a description on how to set up the connection for remote control, see
"Connecting a Controller to the Base Unit"