Instrument Functions
NRP2
122
User Manual 1173.9140.02 ─ 07
Define calculation of measurement result:
Read result:
4.5.4.1
Measurement results
The measured values for the selected gate and trace (waveform) are indicated in the
info panel below the graph (see also
Average / Peak / Peak/Avg
Average power, peak power and the ratio peak power to average power.
"1 | 2"
Start and stop time of gate.
"
Δ
t"
Duration of gate.
Remote command:
Configuring measurement:
Retrieving results:
4.5.4.2
Controls in gates display mode
Display
Config Gates ...
Open the subdialog to configure the four available gates, see
Pos Scale
Defines if the cursor keys position or scale the waveform.
"Pos"
The cursor keys shift the waveform in horizontal and vertical direction
"Scale"
The cursor keys expand and compress the waveform. When the hori-
zontal scaling is modified, the points (pixels) at the left of the screen
remain unchanged. When the vertical scaling is modified, the neutral
zone depends on the unit selected: the points in the center of the
screen remain unchanged when dBm, dB and dBµV are used. The
points at the bottom of the screen remain unchanged in all other dis-
plays.
[1] [2] [3] [4]
Switches the the function of the cursor keys to altering the borders of a gate.
"[1] [2]"
Shifts the start and stop time.
Displaying traces