Remote Control - Commands
NRP2
279
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>]:RGAMma:PHASe
<phase_angle>
SENSe:RGAMma:PHASe
defines the phase angle (in degrees) of the complex reflection
factor of the load at the signal output.
Parameters:
<phase_angle>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: only the unit degree is accepted
[SENSe<[1]..4>]:SAMPling
<sampling_rate>
If the signal to be measured has modulation sections just above the video bandwidth of
the sensor used, measurement errors might be caused due to aliasing effects. In this
case, the sampling rate of the sensor can be set to a safe lower value (
FREQ2
). How-
ever, the measurement time required to obtain noise-free results is extended com-
pared to the normal sampling rate (
FREQ1
).
Parameters:
<sampling_rate>
FREQ1 | FREQ2
*RST:
depending on sensor
[SENSe<[1]..4>]:SGAMma[:MAGNitude]
<magnitude>
The R&S
NRP2 can perform a measured-value correction taking the complex reflection
coefficient (source gamma) of the signal source into account.
SENS<[1]..4>:SGAM:CORR:STAT ON
must be set. The magnitude of the reflection
coefficient is set with
SENS<[1]..4>:SGAM[:MAGN]
.
Parameters:
<magnitude>
float_value
Range:
depending on sensor
*RST:
depending on sensor
[SENSe<[1]..4>]:SGAMma:PHASe
<phase>
The R&S
NRP2 can perform a measured-value correction taking the complex reflection
coefficient (source gamma) of the signal source into account.
SENS<[1]..4>:SGAM:CORR:STAT ON
must be set. The phase angle of the reflec-
tion coefficient is set with
SENS<[1]..4>:SGAM:PHAS
.
Parameters:
<phase>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: degree
SENSe (Sensor Commands)