![Rohde & Schwarz NRP2 User Manual Download Page 400](http://html.mh-extra.com/html/rohde-and-schwarz/nrp2/nrp2_user-manual_1477858400.webp)
Annex
NRP2
399
User Manual 1173.9140.02 ─ 07
The following example is intended for analysing a GSM/EDGE signal with the first slot
used.
*rst
*cls
sens:trac:meas on
'activate pulse measurements
sens:trac:meas:auto on
'for sensor A.
sense:trace:meas:tran:esam:auto:state on 'enhanced accuracy for stable
'periodic signals
sens:trac:meas:alg HIST
'select an algorithm for the pulse
'detection analysis
meas:xtime? (100), 6 ms, (@1)
'perform a trace measurement
'Check the reference levels for
sens:trac:meas:def:dur:ref?
'- pulse duration
sens:trac:meas:def:tran:lref?
'- start of rising/end of falling edge
sens:trac:meas:def:tran:href?
'- start / end of pulse top
sens:trac:meas:puls:dcyc?
'read duty cycle
sens:trac:meas:puls:dur?
'read pulse duration
sens:trac:meas:puls:per?
'read pulse period
sens:trac:meas:puls:sep?
'read pulse separation
sens:trac:meas:tran:neg:dur?
'read length of falling edge
sens:trac:meas:tran:neg:occ?
'read start time of falling edge
sens:trac:meas:tran:neg:over?
'read overshooting of falling edge
sens:trac:meas:tran:pos:dur?
'read length of rising edge
sens:trac:meas:tran:pos:occ?
'read start time of rising edge
sens:trac:meas:tran:pos:over?
'read overshooting of rising edge
sens:trac:meas:power:min?
'read minimum power in trace
sens:trac:meas:power:max?
'read maximum power in trace
sens:trac:meas:power:avg?
'read average power in trace
sens:trac:meas:power:puls:top?
'read mean power level of pulse top
sens:trac:meas:power:puls:base?
'read mean power level of pulse bottom
syst:err?
'check for errors
A.4.7 Synchronizing Measurements
Some sensors (e.g. R&S
NRP-Z81/85/86) are able to generate signals on the (exter-
nal) trigger bus, when an internal trigger was detected. This allows for synchronized
measurements with multiple sensors.
One sensor has to be configured as trigger master (
:TRIG:SOUR INT; :TRIG:MAST
ON
), while all other sensors are slaves with external trigger source (
:TRIG:SOUR
EXT
).
Programming Measurements