Remote Control - Commands
NRP2
272
User Manual 1173.9140.02 ─ 07
Like
"XPOWer:CCDF"
, except that a PDF is calculated instead
of a CCDF.
"XTIMe:POWer"
A sequence of measurements is performed. The individual mea-
sured values are determined as in the ContAv mode. The length
of an individual measurement is determined from the ratio of the
total time (
SENS:TRAC:TIME
) and the number of individual test
points (
SENS:TRAC:POIN
).
*RST:
depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:APERture
<integration_time>
Determines the integration time for a single measurement in the ContAv mode. To
increase the measurement accuracy, this integration is followed by a second averaging
procedure in a window with a selectable number of values. The filter window is config-
ured with the
[SENSe<[1]..4>]:AVERage
commands.
Parameters:
<integration_time>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>][:POWer][:AVG]:BUFFer:SIZE
<number>
Sets the number of desired values for the buffered ContAv mode.
Parameters:
<number>
int_value
Range:
1...400000
*RST:
depending on sensor
[SENSe<[1]..4>][:POWer][:AVG]:BUFFer:STATe
<state>
Switches on the buffered ContAv mode, after which data blocks rather than single
measured values are then returned. In this mode a higher data rate is achieved than in
the non-buffered ContAv mode.
The number of desired measured values is set with the
SENSe<[1]..4>:POWer:ARR:SIZE
command.
Parameters:
<state>
ON | OFF
*RST:
depending on sensor