Remote Control - Commands
NRP2
333
User Manual 1173.9140.02 ─ 07
TRIGger<[1]..4>:HOLDoff
<holdoff>
TRIGger:ALL:HOLDoff
<holdoff>
Defines a period after a trigger event within which all further trigger events are ignored.
Parameters:
<holdoff>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
TRIGger<[1]..4>:HYSTeresis
<value>
TRIGger:ALL:HYSTeresis
<value>
This command is used to specify how far the signal level has to drop below the trigger
level before a new signal edge can be detected as a trigger event. Thus, this command
can be used to eliminate the effects of noise in the signal on the edge detector of the
trigger system.
Unit:
DB | PCT
Parameters:
<value>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DB
TRIGger<[1]..4>[:IMMediate]
TRIGger:ALL[:IMMediate]
Performs triggering and makes sure that the sensor directly changes from the
WAIT_FOR_TRG state to the MEASURING state irrespective of the selected trigger
source (
TRIG:SOUR
). A trigger delay set with
TRIG:DEL
is ignored but not the auto-
matic delay determined when
TRIG:DEL:AUTO:ON
is set. When the trigger source is
HOLD, a measurement can only be started with TRIG.
Error messages:
-211
"Trigger ignored"
: The sensor is not in the WAIT_FOR_TRG state.
Usage:
Event
TRIGger<[1]..4>:LEVel
<level>
TRIGger:ALL:LEVel
<level>
Determines the power a trigger signal must exceed (
TRIG:SLOP POS
) or fall short of
(
TRIG:SLOP NEG
) before a trigger event is detected. This setting is only used for the
trigger signal source (
TRIG:SOUR INTernal
).
Unit:
DBM | W
TRIGger