Remote Control Fundamentals
NRP2
173
User Manual 1173.9140.02 ─ 07
The remote control command
LLO
also allows the ESC/LOCAL key to be disabled. In
this case, return to manual operation is only possible by remote control or by switching
the R&S
NRP2 off and on again (Power On Reset).
5.5 Automatic Settings when Switching between Local
and Remote
The table below lists the settings automatically performed when switching between the
remote and local modes.
Table 5-1: Automatic Settings when switching between Local and Remote
Local to Remote
Remote to Local
SENS:AVER:TCON REP
SENS:AVER:TCON MOV
-
ABORT:ALL
INIT:ALL:CONT ON
DISP:MESS OFF
DISP:MESS OFF
In each CALCulate block, where % is
set as an output unit (
UNIT:POWer:
RATio DPCT
),
UNIT:POWER:RATio
O
is set.
-
The automatic storage of the instru-
ment settings is deactivated in the
nonvolatile memory.
The automatic storage of the instrument settings is reactivated in
the nonvolatile memory.
-
SYST:SPEED NORMal
SENS:BUFF:STAT OFF
SENS:AVER:STAT ON
SENS:TRACe:AVER:STAT ON
SENS:AVER:COUNt:AUTO OFF
-
ContAv:
TRIG:DEL 0
TRIG:HOLD 0
TRIG:SOUR IMM
TRIG:DEL:AUTO OFF
TRIG:COUNt 1
TRIG:ATR OFF
Burst, timegate, timeslot and
trace:
TRIG:DEL:AUTO OFF
TRIG:COUNt 1
TRIG:ATR OFF
5.6 Communication on Data Lines
The GPIB bus is the most common remote interface for measuring instruments. It has
also been implemented in the R&S
NRP2. The remote controller and the R&S
NRP2
Automatic Settings when Switching between Local and Remote