Annex
NRP2
392
User Manual 1173.9140.02 ─ 07
Trigger level
Power
Time
Trigger event
SENS:TRAC2:OFFS:TIME
SENS:TRAC2:TIME
Gate 5
SENS:TRAC2:MID:TIME
SENS:TRAC2:MID:OFFS:TIME
Fig. 1-9: Measurement of a GSM slot with Fence(Mid)-Exclude domain
Level, hysteresis and dropout time
With internal and external triggering (
TRIG:SOUR INT | EXT
), a trigger event is rec-
ognized when the level of a trigger signal exceeds a specific value. This value is the
trigger level and can be set with
TRIG:LEV
. With internal triggering, the end of the
measurement is recognized when the power level falls below the trigger level.
This is not desirable with modulated signals or short interruptions and also when the
level just falls slightly below the trigger level:
●
SENS:BURS:DTOL
is used to define a time span during which the power to be
measured in the "BurstAv" mode must remain below the trigger level so that the
end of the burst can be detected.
●
TRIG:HYST
is used to specify a value in dB by which the signal must fall below the
trigger level so that the end of a burst can be detected.
A.4.4.3
Configuration of sensor settings
The sensor settings are controlled by the
SENSe
command system. The sensors tell
the basic unit which commands they understand and which parameters are allowed.
As a result, the commands described below cannot be used for some sensors.
Selecting a measurement mode
The most important sensor-dependent setting is the selection of a measurement mode:
ContAv mode:
SENS:FUNC "POW:AVG"
Continuous measurement with only one integration time specified (
SENS:APER
), within
which the power is integrated as a measurement result.
Programming Measurements