Remote Control - Commands
NRP2
278
User Manual 1173.9140.02 ─ 07
[SENSe<[1]..4>][:POWer]:TSLot:[AVG]:WIDTh
<time>
Sets the length of the timeslot in the Timeslot mode (see figure below).
Trigger level
Power
Time
Trigger event
SENS:TSLOT:WIDTH
SENS:TSLOT:MID:OFFS:TIME
SENS:TSLOT:MID:TIME
T'Slot 1
T'Slot 2
T'Slot 3
SENS:TSLOT:COUNT 3
Fig. 6-13: Timeslot parameters
Parameters:
<time>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>]:RGAMma[:MAGNitude]
<magnitude>
The
SENSe:RGAMma
command group compensates for the reflection of the load at the
signal output. Such a compensation is necessary if the VSWR exceeds a value of 1.05.
Otherwise data sheet specifications for absolute accuracy cannot be met reliably.
To compensate for reflection, the magnitude of the reflection coefficient of the load with
the
SENSe:RGAMma:MAGNitude
setting and the phase angle with the
SENSe:RGAMma:PHASe
setting must be transferred. To deactivate reflection compen-
sation,
SENSe:RGAMma:MAGNitude 0
must be set. Unlike with mismatch correction,
there is no special activation/deactivation command. Reflection compensation should
remain deactivated if the values measured for the load reflection coefficient are disput-
able.
Parameters:
<magnitude>
float_value
Range:
depending on sensor
*RST:
depending on sensor
SENSe (Sensor Commands)