Remote Control - Commands
NRP2
276
User Manual 1173.9140.02 ─ 07
Trigger level
Power
Time
Trigger event
Delayed
trigger point
Trigger
delay
Gate 1
Gate 4
Gate 3
Gate 2
SENS:TGAT1:OFFS:TIME
SENS:TGAT1:TIME
SENS:TGAT2:OFFS:TIME
SENS:TGAT2:TIME
Fig. 6-12: Timing parameters for the Timegate mode
Parameters:
<time>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
[SENSe<[1]..4>][:POWer]:TGATe:SELect
<time_gate>
Selects one of the four gates in which the average power is measured if the Timegate
mode is selected.
Parameters:
<tgate>
1 | 2 | 3 | 4
*RST:
1
[SENSe<[1]..4>][:POWer]:TGATe<[1]..4>:TIME
<time>
Is used for configuring the Timegate mode. It specifies the duration of one of the four
gates (sweeps) relative to the delayed trigger point. In manual operation, the start and
duration of the gates can also be defined graphically if the sensor is operated in the
Trace mode.
Parameters:
<time>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: S
SENSe (Sensor Commands)