Remote Control - Commands
NRP2
265
User Manual 1173.9140.02 ─ 07
However,
SENS<[1]..4>:AVER:COUN:AUTO:RES
does not affect the
DISPlay
command. The parameters of the two commands are different but have the same
meaning.
SENS<[1]..4>:AVER:COUN:AUTO:RES
1
2
3
4
DISP<[1]..4>:RES
1
0.1
0.01
0.001
Parameters:
<<resolution>>
1 | 2 | 3 | 4
Range:
depending on sensor
*RST:
depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:SLOT
<slot>
Sets a timeslot whose measured value is used to automatically determine the filter
length.
Parameters:
<slot>
int_value
Range:
depending on sensor
*RST:
depending on sensor
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:NSRatio
<nsratio>
Indicates the maximum noise ratio in the measurement result. Achieving the same
result by filter length setting is attempted only if
SENS<[1]..4>:AVER:COUN:AUTO ON
and
SENS<[1]..4>:AVER:COUN:AUTO:TYPE NSR
are set.
Unit:
DB | PCT
Parameters:
<nsratio>
float_value
Range:
depending on sensor
*RST:
depending on sensor
Default unit: DB
[SENSe<[1]..4>]:AVERage:COUNt:AUTO:TYPE
<mode>
Selects a method by which the automatic filter length switchover can operate:
SENSe (Sensor Commands)