Remote Control - Commands
NRP2
227
User Manual 1173.9140.02 ─ 07
Note:
Every measured value displayed in the R&S
NRP2 (also every point of a trace) is aver-
aged in the sensor via a series of samples. If a
RANDom
feed is selected, a random
value is taken from these samples and forwarded to the base unit as a measured
value.
PEAK
is the maximum of all samples in the measurement interval.
Suffix:
<[1]|2]>
.
Parameters:
<feed>
string
see
*RST:
"POWer:AVERage"
CALCulate<[1]..8>:LIMit:BEEP
<beep>
Switches the acoustic warning for limit violations on or off.
Parameters:
<beep>
ON | OFF
*RST:
OFF
Usage:
Setting only
CALCulate<[1]..8>:LIMit:CLEar[:IMMediate]
Resets the limit monitoring state and the internal counter for limit violations.
Usage:
Event
CALCulate<[1]..8>:LIMit:CLEar:AUTO
<auto>
Automatically resets the limit monitoring state and the internal counter for limit viola-
tions if any of the following events occurs:
●
INITiate[:IMMediate]
is executed
●
INITiate:CONTinuous ON
is executed
●
MEASure?
or
MEASure:...?
query is executed.
●
READ?
or
READ:...?
query is executed.
CALCulate (Configuration of Calculate Blocks)