Instrument Functions
NRP2
84
User Manual 1173.9140.02 ─ 07
Fig. 4-18: Range dialog box
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Auto
Turns the "Auto" range function on/off.
Remote command:
[SENSe<[1]..4>][:POWer][:AVG]:RANGe:AUTO
Path 1 2 3
Selects a measurement path when the "Auto" range function is off.
Remote command:
[SENSe<[1]..4>][:POWer][:AVG]:RANGe
User def'd transition
Activates a shift of the transition region(s).
Remote command:
[SENSe<[1]..4>][:POWer][:AVG]:RANGe:CLEVel
Offset
Negative dB value as a measure for the shift of the transition region(s).
Modifications are immediately effective after each keystroke (see
Remote command:
[SENSe<[1]..4>][:POWer][:AVG]:RANGe:CLEVel
4.2.5 Effect of the RF source
Information related to the reflection coefficient
Γ
of the signal source is provided in the
"
Γ
Correction" dialog box (gamma correction).
Γ
Correction is used to increase mea-
surement accuracy by correcting the effect of source mismatch on the measured value.
Data acquisition and parameters