Remote Control - Commands
NRP2
307
User Manual 1173.9140.02 ─ 07
Bit No.
Meaning
9
POST Failure
The built-in test of the R&S
NRP2 carried out automatically upon power-up has gener-
ated an error.
10 to 13
Not used
14
Warning
15
Bit 15 will never be used.
6.13.9 Standard Event Status and Enable Register (ESR, ESE)
The ESR is already defined in the IEEE 488.2 standard. It is comparable to the EVENt
register of an SCPI register. The Standard Event Status Register can be read out by
the query
*ESR?
.
The ESE forms the associated ENABle register. It can be set by the command
*ESE
and read out by the query
*ESE?
.
Table 6-21: Meaning of bits used in the Standard Event Status Register
Bit No.
Meaning
0
Operation Complete
When the
*OPC
command is received, this bit is set if all previous commands have been
executed.
1
Not used
2
Query Error
This bit is set in either of the two following cases: the controller wants to read data from the
device but has not sent a query, or it sends new commands to the device before it retrieves
existing requested data. A frequent cause is a faulty query which cannot be executed.
3
Device-dependent Error
This bit is set if a device-dependent error occurs. An error message with a number
between -300 and -399 or a positive error number denoting the error in greater detail will
be entered in the error queue.
4
Execution Error
This bit is set if the syntax of a received command is correct but the command cannot be
executed due to various marginal conditions. An error message with a number between
-200 and -300 denoting the error in greater detail will be entered in the error queue.
5
Command Error
This bit is set if an undefined command or a command with incorrect syntax is received. An
error message with a number between -100 and -200 denoting the error in greater detail
will be entered in the error queue.
6
User Request
This bit is set upon pressing the [LOCAL] key, i.e. when the device is switched to manual
control.
7
Power On
This bit is set when the device is switched on.