Annex
NRP2
396
User Manual 1173.9140.02 ─ 07
SENS:CORR:FDOF?
'queries correction value used
INIT
FETCh?
This example assumes that a table called "Table 1" is available.
Duty cycle
If the duty cycle of a pulsed signal (
SENS:DCYC
) is reported to the R&S
NRP2, the
R&S
NRP2 delivers the average power in the pulse.
Example
*RST
SENS:CORR:DCYC:STAT ON
'duty-cycle correction
SENS:CORR:DCYC 30 PCT
'duty cycle of 30%
INIT
FETCh?
Reflection coefficient of the source (source gamma,
Γ
)
Reflections at the power source may distort the result. If the reflection coefficient of the
source is known, it can be specified for the measured-value correction via command
path
SENS:SGAM
.
Example
*RST
SENS:SGAM:CORR:STAT ON
'considers the reflection coefficient of the source
SENS:SGAM:MAGN 0.01
'magnitude of the reflection coefficient
SENS:SGAM:PHAS 179
'phase angle of the reflection coefficient
INIT
FETCh?
A.4.4.4
Configuration of CALCulate blocks
The function of the
CALCulate
blocks has already been explained. Its most important
task is the calculation and provision of measurement results.
Calculation function and relative measurement
The calculation functions mentioned in connection with high-level commands are inter-
nally implemented via commands
CALC:MATH
and
CALC:REL:STAT
.
Keyword for selecting a calculation
function with high-level commands
Corresponding low-level commands (n,m=1,2,3,4)
:DIFF
(Difference)
CALC:MATH "(SENSn-SENSm)"
;
REL:STAT OFF
:DIFF:REL
(Difference with refer-
ence value)
CALC:MATH "(SENSn-SENSm)"
;
REL:STAT ON
:RAT
(Ratio)
CALC:MATH "(SENSn/SENSm)"
;
REL:STAT OFF