
Instrument Functions
NRP2
79
User Manual 1173.9140.02 ─ 07
Fig. 4-15: Table dialog box
The table editor displays the name of the table in its title bar, a two-column list with fre-
quency and offset value pairs and provides softkeys for data management.
Edit
Activates the edit mode in the selected field.
The entry can be confirmed with the
↵
MENU or aborted with the ESC/LOCAL key.
Afterwards, the table editor is again in the selection mode.
Insert
Inserts a new row containing the values of the selected row.
The maximum table size provides 80 rows.
Delete
Deletes the marked row
Sort
Sorts the rows in the table so that the frequencies in the rows are in ascending order.
Name
Renames the selected table to identify it more easily. Use of the character editor is
described in
chapter 3.5.9, "Digit and letter panels"
Remote command:
4.2.3 Filter / averaging
The averaging filter is used to reduce fluctuations in the measurement result to the
extent wanted. Such fluctuations can be caused by inherent noise of the measuring
instrument, modulation of the measurement signal or beats from the superposition of
adjacent carriers. A more stable display has to be traded off against longer measure-
ment times.
Data acquisition and parameters
北京海洋兴业科技股份有限公司(证券代码:839145)
电话:010-62176775
网址:www.hyxyyq.com