Annex
NRP2
395
User Manual 1173.9140.02 ─ 07
INIT
FETCh?
All intermediate pushed values delivered by the first filter stage are stored in the filter.
The earliest measured value is ousted out of the filter, since the filter has only a limited
length.
SENS:AVER:TCON MOV | REP
is used to set when the filter algorithm is to be
applied to the filter. With the setting
MOVing
, a new measure value is calculated for
each new intermediate measured value. This is the default setting in the
LOCAL
mode.
SENS:AVER:TCON REPeat
is set in the
REMOTE
mode. In this particular case, new
measured values are calculated when all intermediate measured values have been
replaced by new ones.
Changing from
SENS:AVER:TCON MOV
to
REP
in the
LOCAL-REMOTE
transition cau-
ses the update rate of measured values to be reduced, since the filter must be com-
pletely filled for each new measured value.
Measured-value corrections
The R&S
NRP2 offers different options for correcting measured values in the instru-
ment. These options all require an existing knowledge of the test setup or the time
structure of the signal. Offsets in the result can be corrected globally and depending on
the frequency, and it is possible to consider the duty cycle of a signal and the influence
of the complex reflection coefficient (
Γ
) of the power source.
Global offset correction
All measured values are multiplied by a correction factor (logarithmically added).
*RST
SENS:CORR:OFFS:STAT ON
'global offset correction ON
SENS:CORR:OFFS 20 DB
'corrects result by +20 dB
INIT
FETCh?
Frequency-dependent offset correction
Two-column tables can be entered in which frequencies and correction values can be
stored. A frequency-dependent correction value is determined by linear interpolation in
the units Hz and dB using the signal frequency (defined with
SENS:FREQ
). The interpo-
lated correction value can be queried with
SENS:FDOF?
for checking purposes.
Example
*RST
MEM:TABL:MOVE "Table 1" "Splitter" 'renames first table
MEM:TABL:SEL "Splitter"
'selects table
MEM:TABL:CLEar
'deletes selected table
MEM:TABL:FREQ 0,1e4,5e4,1e5,1e9
'interpolation points on the frequency axis
MEM:TABL:GAIN 3.1,3.1,3.0,2.9,2.9
'associated offset values
SENS:FREQ 900 MHZ
'signal frequency
SENS:CORR:FDOT:STAT ON
'activates frequency-dependent offset correction
SENS:CORR:FDOT "Splitter"
'selects an offset table